|
This supplemental report is being submitted to provide the results of the final investigation and additional information.Updated fields: b5, d9, h2, h3, h6, h11.The device was returned to olympus for inspection and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: dented outer tube, broken light guide prong and image view out of field.A root cause could not be identified.Based on the results of the investigation, the most probable cause of the connection problem was due to broken light guide prong, the most probable cause of out of view field image was bent outer tube.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|