Model Number 7300TFX27
Medical Device Problem Code Calcified (1077)
Health Effect - Clinical Code Heart Failure/Congestive Heart Failure (4446)
Date of Event 02/04/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H11.Additional narrative.Surgical/percutaneous intervention is indicated or performed, or harm occurred due to the device, or there is a device malfunction that could cause or contribute to a serious injury.This event is considered a serious injury.The device was not returned for evaluation, as it was discarded.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.
 
Event or Problem Description
It was reported that a 7300tfx 27mm mitral valve was explanted and replaced with a non-ew valve after an implant duration of 8 years, 10 months due to severe stenosis and heavy calcification.The patient presented with acute on chronic hf with preserved ef.Per medical records, intraoperative inspection of mv revealed heavy calcification.The surgeon resected the valve and replaced it with a non-ew porcine bioprostheses.The patient tolerated the procedure well and was taken to icu in stable condition.The patient discharged on pod#6.
 
Additional Manufacturer Narrative
H11.Additional narrative updated h6 svd is a logical and expected consequence of the chemical, mechanical, and immunological processes associated with bhv implantation.The common endpoint of all these factors is calcification and degradation.Svd is an acquired condition, intrinsic to bioprosthetic valves, characterized by deterioration of the leaflets or supporting structures, leading to thickening, calcification, tearing, or disruption of the prosthetic valve materials.These changes result in valvular dysfunction manifesting as stenosis and/or regurgitation with a consequent drop in hemodynamic efficacy of the valve.Despite the advancements in the heart valve industry, svd commonly begins 7- 8 years after implantation, with a considerably increased rate in patients with pertinent comorbidities and/or an implant duration of greater than 10 years.Prior investigations and extensive literature review have shown that svd is predominantly related to patient factors and implant duration rather than to manufacturing issues.Events related to device design, manufacturing, or use error tend to manifest at an earlier implant duration.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.Based on the information available, the most likely root cause is patient factors, including an implant duration of 7 or more years and patient history of dyslipidemia, hyperlipidemia, and smoking.The device history record (dhr) was reviewed and showed that this device met all manufacturing and sterilization specifications for product release prior to distribution.No issues were identified that would have impacted this event.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL EASE PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer Contact
jason reves
one edwards way
mle fl2
irvine, CA 92614
9492503859
MDR Report Key24590932
Report Number2015691-2026-13180
Device Sequence Number10114873
Product Code DYE
UDI-Device Identifier00690103186260
UDI-Public(01)00690103186260(17)181202(11)141203
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/02/2018
Device Model Number7300TFX27
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2026
Supplement Date Received by Manufacturer04/07/2026
Initial Report FDA Received Date03/13/2026
Supplement Report FDA Received Date04/09/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/03/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization; Life Threatening;
Patient Age65 YR
Patient SexMale
-
-