| Model Number |
TSGC0801 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is not returning for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a patient presented with grade 5 functional tricuspid regurgitation (tr) for a triclip procedure.During the procedure, the triclip was successfully placed on the leaflets and deployment steps were performed.During lock line removal the clip opened to approximately 45 degrees.Troubleshooting was preformed and the clip arms were closed.The clip remained close while the rest of the lock line was removed.The procedure was discontinued.The final tr was grade 2.After the procedure, it was difficult to remove the steerable guide catheter (sgc) from the stabilizer.The sgc was removed from the patient, still attached to the stabilizer.Troubleshooting was preformed, after significant force the sgc was removed from the stabilizer.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to the reported difficult to remove.Additionally, a review of the complaint history identified similar complaints reported to this lot, and this device appears to be related to a potential product quality issue.The investigation determined that the reported difficult to remove may be related to a potential product quality issue.This complaint falls within the scope of an exception, as the complaint description and device code match the specific issue described in the exception.The issue is being addressed per internal operating procedures.Abbott will continue to trend the performance of these devices.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|