• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL TRICLIP¿ G5 STEERABLE GUIDE CATHETER; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL TRICLIP¿ G5 STEERABLE GUIDE CATHETER; CATHETER, PERCUTANEOUS Back to Search Results
Model Number TSGC0801
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/18/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not returning for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2026, a patient presented with grade 5 functional tricuspid regurgitation (tr) for a triclip procedure.During the procedure, the triclip was successfully placed on the leaflets and deployment steps were performed.During lock line removal the clip opened to approximately 45 degrees.Troubleshooting was preformed and the clip arms were closed.The clip remained close while the rest of the lock line was removed.The procedure was discontinued.The final tr was grade 2.After the procedure, it was difficult to remove the steerable guide catheter (sgc) from the stabilizer.The sgc was removed from the patient, still attached to the stabilizer.Troubleshooting was preformed, after significant force the sgc was removed from the stabilizer.There were no adverse patient effects or clinically significant delays reported.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to the reported difficult to remove.Additionally, a review of the complaint history identified similar complaints reported to this lot, and this device appears to be related to a potential product quality issue.The investigation determined that the reported difficult to remove may be related to a potential product quality issue.This complaint falls within the scope of an exception, as the complaint description and device code match the specific issue described in the exception.The issue is being addressed per internal operating procedures.Abbott will continue to trend the performance of these devices.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRICLIP¿ G5 STEERABLE GUIDE CATHETER
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key24591263
Report Number2135147-2026-01625
Device Sequence Number13160242
Product Code DQY
UDI-Device Identifier05415067050724
UDI-Public(01)05415067050724(17)270430(10)51031R1081
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K243224
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTSGC0801
Device Catalogue NumberTSGC0801
Device Lot Number51031R1081
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/18/2026
Supplement Date Received by Manufacturer03/18/2026
Initial Report FDA Received Date03/13/2026
Supplement Report FDA Received Date03/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/31/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
STABILIZER
Patient Age87 YR
Patient SexFemale
Patient Weight64 KG
Patient EthnicityNon Hispanic
-
-