| Catalog Number |
5513-001 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Further information regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event is a known potential adverse event addressed in the product labeling.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a xen®45 gts "unable to move smoothly and stuck when trying to deploy" in left eye.Surgery was completed with backup device.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a xen®45 gts "unable to move smoothly and stuck when trying to deploy" in left eye.Surgery was completed with backup device.
|
| |
|
Additional Manufacturer Narrative
|
|
Device analysis:
one xen 45 glaucoma treatment system (gts) injector was received.The slider position appears to be between the start and the end position, however, since the slider knob was not properly positioned within the track of the cam, proper determination of slider position cannot be made.A stent was not observed within the returned packaging.It is unknown whether the stent is within the injector.The complainant did not report that the slider knob was mispositioned.The condition of the mispositioned slider knob tab is likely due to complainant handling.No further evaluation of this damage is required.Slider resistance and injector was unable to verify since the injector was received with the slider knob mispositioned.It is unknown if internal components are properly aligned; therefore, functional evaluation cannot be performed.The complainant did not report that the slider knob was mispositioned.The condition of the mispositioned slider knob is likely due to complainant handling.No further evaluation of this damage is required.Additional, changed, and/or corrected data: d9, h3, h6.
|
| |
|