| Catalog Number |
826850 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/19/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
2 of 3 reports.Other mfg report numbers: 3006697299-2026-00024, 3014334038-2026-00025.A facility reported: the monitored icp pressure readings getting lower and lower until it disappears from the monitor screen all together.It was re-calibrated and functionally checked & tested using an in-house test sensor and all seems to be working correctly.Patient pressures reading 28.Monitor spontaneously showed varying pressures from minus 20 to plus10.Message: "possible input failure.Verify sensor,cable, and lead connections".The earthing wire was attached to the patient.Then the message: "patient monitor connection error, reconnect monitor again"; patient was never disconnected.Then monitor showed no pressure readings.Moved patients head and pressures reappeared.Then numerical pressures disappeared again.Monitor re-zeroed and turned off and on.Message: "possible input failure verify sensors,cables and lead connections" reappeared.There was always a continous error message throughout the incident.Evd transduced and showed icps 18.Medic notified and referral to neurosurgeons, evd had been transduced earlier to check pressure and at that point both the cerelink and evd pressures were 18.The monitor records in the errors logs it showing the following errors "err: 0x0001 sensor warning: mean icp value exceeds the valid range" followed by a low minus value which over a minute or two increases from -60.69 to a value of ¿ 318.25.The log then starts to display multiple "err: 0x1019 param1: 0000 param2: 0000 error type: 0001¿ errors.Unit cleaned/inspected ok.Unit powered up ; re-caled ; cal checked ok.Within spec.Unit functionally checked ; tested ok.Nff.Units history event logs uploaded ; inspected.As seen before the monitored pressure readings start to display high minus values ; then displays in the log "err: 0x0001 sensor warning: mean icp value exceeds the valid range" followed by a minus value.Then displays multiple "err: 0x1019 param1: 0000 param2: 0000 error type: 0001.Suspect repeat problem with faulty reading from probe sensor.Additional information received: date of event?19/02/2026 was the patient lead (electrode of extension cable) always connected to the patient? yes implant location - subdural parenchyma.Did the issue lead to any patient injury / complications or death? no did the issue lead to a critical delay in patient care? no what was the user doing (e.G., powering unit, zeroing sensor, setting alarm, downloading data)? nothing.What was happening with patient (e.G., transport, mri, ct)? nothing ¿ in icu what did medical team do to solve the issue? new probe inserted.If applicable: can you please confirm if the replacement with new sensor was successful?yes.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h11.The cerelink sensor was returned for evaluation:
dhr - lot 7588269 sn (b)(6), conformed to the specifications when released to stock.Failure analysis - the issue of the complaint was confirmed:
- icp express = no transducer detected.- cerelink = sensor failure detected.- there is a kink in the catheter located 2.2 cm from the distal tip.- suspected foreign matter ingress in the catheter approximately 3.1 cm from the distal tip.- multiple small cuts are present on the catheter nylon, located approximately 6.4 cm from the distal tip.- internal wires broken.- no further testing possible.Root cause - the complaint was confirmed by the supplier.The root cause of the issue reported could be due to mishandling of the device.
|
| |
|