| Model Number |
CI-1600-05 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Codes |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Event or Problem Description
|
|
A review of the test data indicated impedance issues.Programming adjustments were made, however, the issue did not resolve.Revision surgery scheduled.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information section: b.3, d.6b, d.9 & h.6.Advanced bionics considers the investigation into this reportable event as closed.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.The external visual inspection revealed the electrode was severed.This is believed to have occurred during revision surgery.The device passed photographic imaging inspection.System lock was verified.The condition of the electrode prevented an electrical test from being performed.The device passed some of the electrical tests performed.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was implemented.This version of the hires ultra device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Search Alerts/Recalls
|