| Model Number |
V+O+C+S+N, ENGLISH |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H6: the device was evaluated by ventec where the reported issue of it rebooting by itself was confirmed.Ventec opened the device in order to perform an internal inspection and observed that the device's cough assist valve had a torn poppet ring.Ventec replaced the cough assist valve to resolve the reported issue.Proper device operation was then confirmed through functional and performance testing.The investigation determined that the root cause of the reported issue was the cough assist valve's poppet ring was torn.
|
| |
|
Event or Problem Description
|
|
It was reported to ventec that the device continuously reboots by itself.There were no reports of patient involvement associated with the reported issue.
|
| |
|