C.R. BARD INC. (COVINGTON) -1018233 PUREWICK FEMALE EXTERNAL CATHETER Back to Search Results
Catalog Number PWFX30
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Pressure Sore/Ulcer (2326)
Date of Event 03/09/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
Patient said she has a decubitus ulcer from using pw in the hospital.It is unclear if the lot number was requested.No medical intervention was reported.No additional information provided.
 
Additional Manufacturer Narrative
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.The reported event is addressed within the labeling as it states: warnings: do not use the purewick female external catheter with bedpan or any material that does not allow for sufficient airflow.Discontinue use if an allergic reaction occurs.Precautions: not recommended for patients who are: agitated, combative, or uncooperative and might remove the purewick female external catheter.Having frequent episodes of bowel incontinence without a fecal management system in place.Experiencing skin irritation or breakdown at the site.Experiencing moderate/heavy menstruation and cannot use a tampon.Do not use barrier cream on the perineum when using the purewick female external catheter.Barrier cream may impede suction.Proceed with caution in patients who have undergone recent surgery of the external urogenital tract.Always assess skin for compromise and perform perineal care prior to placement of a new purewick female external catheter.Maintain suction until the purewick female external catheter is fully removed from the patient to avoid urine backflow.Recommendations: perform each step with clean technique.In the home setting, wash hands thoroughly before device placement.Prior to connecting the purewick female external catheter to hospital wall suction tubing, verify suction function by covering the open end of the suction tubing with one hand and observing the pressure dial.If the pressure does not increase when the line is covered, verify that the tubing is secured, connected, and not kinked.Ensure the purewick female external catheter remains in the correct position after turning the patient.Remove the purewick female external catheter prior to ambulation.Properly placing the purewick female external catheter snugly between the labia and gluteus holds the purewick female external catheter in place for most patients.Mesh underwear may be useful for securing the purewick female external catheter for some patients.Assess device placement and patient¿s skin at least every 2 hours.Replace the purewick female external catheter every 8 to 12 hours or when soiled with feces or blood.Change suction tubing per hospital protocol or at least every thirty (30) days.A dhr could not be performed since no lot number was provided.No additional actions can be taken at this time.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the patient said she has a decubitus ulcer from using pw in the hospital.It is unclear if the lot number was requested.No medical intervention was reported.No additional information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUREWICK FEMALE EXTERNAL CATHETER
Common Device Name
PUREWICK
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key24596314
Report Number1018233-2026-01754
Device Sequence Number8688593
Product Code NZU
UDI-Device Identifier00801741189050
UDI-Public(01)00801741189050
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue NumberPWFX30
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/09/2026
Supplement Date Received by Manufacturer04/02/2026
Initial Report FDA Received Date03/13/2026
Supplement Report FDA Received Date04/03/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-