|
No product has been returned to the investigation site for evaluation; therefore, the condition of the product could not be verified.The most common causes of (tss) toxic anterior segment syndrome are identified as follows in order of prevalence: inadequate flushing, irrigation/aspiration handpieces and cannulated equipment, use of enzymatic cleaners and detergents, use of reusable cannulas, inadequate cleaning of instruments, use of preserved epinephrine, reuse of single use devices, use of tap water with no sterile water final rinse, inadequate personnel or trays to allow proper preparation of instruments, no immediate cleaning allowing ophthalmic viscoelastic device (ovd) and surgical solutions to dry on instruments, use of preserved medicines in the eye, reuse of tubing for flushing, latex bulbs for irrigation, not training, no terminal sterilization, instruments stored on towels, touching or patient contact areas of instruments with gloved hands, off-label use of lidocaine gel, poor instrument maintenance, autoclave residue, rust, particulates, lint, use of powdered gloves, additives added to balanced salt solution against directions for use (dfu) improper use of prep solutions, detergents and cleaners, failure to follow manufacturer¿s directions for use, including no air flush, use of unapproved enzymatic cleaners, use of postoperative ointment in clear corneal cases, povidone-iodine placed in the eye at the end of procedures, incorrect concentration of detergents and enzymatic cleaners.The phacoemulsification systems are closed systems.They are operated with a sterile single use consumable cassette which is designed to isolate the patient fluid path from the console itself.Any surgical instrumentation that would come into contact with the patient would be cleaned and autoclaved by the user prior to surgery, per standard industry practices and company directions for use (dfu).The proper cleaning and sterilization of ophthalmic surgical instruments can help prevent the occurrence of tass.These findings continue to validate the need to follow the recommendations detailed in the dfus, recommended practices.There is no evidence that the consumable cassette contributed to the reported event.The investigation conducted a non-conformance review of the reported lot number.No deviations that could have contributed to this event were identified and all corresponding production release specifications defined in the device master record were met.Based on the available information and no materials returned for product evaluation, the root cause of this tass event is inconclusive.No further investigative or manufacturing actions are warranted at this time as there was no verification of product condition and no product-related factors contributing to the event.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|