Catalog Number 03.037.016
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: h3, h6: product investigation summary: the product was returned to depuy synthes for evaluation.Visual inspection of the returned device found distal end of the device is nicked.An interaction was made using the guide sleeve yell and buttress/compr-nut and found that devices are unable to be disassembled due to guide sleeve yell being nicked at the threads.The observed condition of the device is consistent with a component failure that was caused by exposure to unintended forces like overtightening the device while assemblance; or by assembling the device in a misaligned position.A dimensional inspection was unable to be performed due to the post-manufacturing damage.A functional test was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the buttress/compr-nut would have contributed to the complained issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is required.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H3, h4, h6: device history record (dhr) review conducted: part: 03.037.016.Lot: 9841605.Manufacturing site: werk hagendorf.Supplier: na.Release to warehouse date: 04 may 2016.Expiration date: na.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that during pre-operative inspection it was identified that the buttress/compression-nut is stuck on the sleeve.There was no patient impact.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d10.H3, h6: the product was returned to depuy synthes for evaluation.Visual inspection of the returned device found distal end of the device is nicked.An interaction was made using the guide sleeve yell and buttress/compr-nut and found that devices are unable to be disassembled due to guide sleeve yell being nicked at the threads.The observed condition of the device is consistent with a component failure that was caused by exposure to unintended forces like overtightening the device while assemblance; or by assembling the device in a misaligned position.A dimensional inspection was unable to be performed due to the post-manufacturing damage.A functional test was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the buttress/compr-nut would have contributed to the complained issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is required.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: part: 03.037.016, lot: 9841605, manufacturing site: werk hagendorf, supplier:na, release to warehouse date:04 may 2016, expiration date; na.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BUTTRESS/COMPRESSION NUT
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
SYNTHES HAEGENDORF GMBH
im bifang 6
haegendorf 4614
SZ   4614
Manufacturer Contact
elba bello
1302 wrights lane east
west chester, PA 19380
9083863534
MDR Report Key24597805
Report Number8030965-2026-02492
Device Sequence Number15843219
Product Code LXH
UDI-Device Identifier10886982070326
UDI-Public(01)10886982070326(10)9841605
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.037.016
Device Lot Number9841605
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/24/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/03/2026
Supplement Date Received by Manufacturer03/31/2026
Initial Report FDA Received Date03/13/2026
Supplement Report FDA Received Date04/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/04/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DRILLSL Ø3.2 F/NEUTRAL POSITION F/PERIAR; GUIDE SLEEVE YELL; SCRDRIVER SHAFT 1.3/1.5 T4 SELF-HOLDING; SCRDRIVER SHAFT 2.4 SHORT SELF-HOLDING F; SCRDRIVER SHAFT 3.5 T15 SELF-HOLDING F/A; SCRDRIVER SHAFT SELF-HOLDING L165 F/STAR; SCRDRIVER SHAFT SELF-HOLDING L165 F/STAR; SCRDRIVER SHAFT T25 L100; SCRDRIVER SHAFT T8 SELF-HOLDING; SCRDRIVER SHAFT T8 SELF-HOLDING; SCRDRIVER SHAFT T8 SELF-HOLDING; SCRDRIVER-HEX Ø3.5 COMBINED W/STARDRIVE; SD/HD SCRWDRVR SFT T25 3.5HEX/SF-RTN 280