|
This report is being submitted to provide additional information.Updated: b4, b5, d2b, g1, g3, g6, h1, h2, h3, h6, and h11 visual examination of the provided pictures identified the peg of the patellar implant has been fractured off from the body of the implant.As the device was not returned, further evaluations could be provided.Medical records were not provided.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.This complaint is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|