| Model Number |
P1170B |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Baxter technical support provided the account the part number of the external alarm that needs to be replaced to resolve the issue.The account will repair the bed themselves.Based on this information, no further action is required.Per the baxter service manual the bed should be subject to an effective maintenance program.An annual service of the bed is advised in order to maintain its characteristics and performance.Brake casters should be checked for cuts, wear and quality of tread, etc.And replaced when necessary.Check the brakes to see whether the bed moves when the brake pedals are pressed and repair as necessary.A search of the baxter maintenance records did not show baxter performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.No further information is available on the repair of the bed at this time.If any additional relevant information is identified following completion of the repair, the additional relevant information will be submitted in a supplemental.
|
| |
|
Event or Problem Description
|
|
On 13-feb-2026, a other health care professional contacted technical service to report that care assist ca100 (product code p1170b0000020, serial number (b)(6)), had brakes not holding.There was no patient/user injury, medical intervention, symptom, or adverse event reported.
|
| |
|