• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 15750 ALTON PKWY RADIUS VSM; MONITOR, PHYSIOLOGICAL, PATIENT (WITH ARRHYTHMIA DETECTION OR ALARMS)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 15750 ALTON PKWY RADIUS VSM; MONITOR, PHYSIOLOGICAL, PATIENT (WITH ARRHYTHMIA DETECTION OR ALARMS) Back to Search Results
Model Number 9037
Medical Device Problem Codes No Display/Image (1183); Sensing Intermittently (1558)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/13/2026
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported the disappearance "of temperature and nibp measurement" on the radius vsms.Measurements from the vsm showed up blank on the psn and there was no error nor alarm triggered.There was no patient impact or consequence reported.
 
Additional Manufacturer Narrative
Other text: masimo has reached out to the customer to request the return of the device.The device has not been returned to masimo to allow an analysis to be performed.If the device is returned for evaluation or new information is obtained, a follow-up report will be submitted.Health effect impact code: no adverse event, no patient impact.D11.Concomitant medical products: the customer reported electrode leads and a blood pressure module were used in conjunction with the suspect medical device.However, the specific lot number of the electrode leads used could not be determined.Therefore, both electrode lead lot numbers were included in as concomitant products in this mdr.E1.Initial reporter zip code exceeded allowable field length, initial reporter zip code is as follows: (b)(6).E1.Initial reporter phone number exceeded allowable field length, initial reporter phone number is as follows: (b)(6).
 
Event or Problem Description
The customer reported the disappearance "of temperature and nibp measurement" on the radius vsms.Measurements from the vsm showed up blank on the psn and there was no error nor alarm triggered.There was no patient impact or consequence reported.
 
Additional Manufacturer Narrative
Other, other text: the returned radius vsm was evaluated.The device was able to obtain measurements for all enabled parameters.The radius vsm was able to connect to the safetynet system and both the vsm and the safetynet system would audible and visually alarm when the alarm limits were breached.No product performance issues related to temperature or blood pressure readings were identified during testing.The device was determined to be functioning as designed.Health effect impact code: no adverse event, no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADIUS VSM
Common Device Name
MONITOR, PHYSIOLOGICAL, PATIENT (WITH ARRHYTHMIA DETECTION OR ALARMS)
Manufacturer (Section D)
MASIMO - 15750 ALTON PKWY
15750 alton pkwy
irvine CA 92618
Manufacturer (Section G)
MASIMO - MEXICALI
industrial vallera de mexicali calzada del oro, no.2001
mexicali, baja california 21600
MX   21600
Manufacturer Contact
lusine boyadzhyan
15750 alton pkwy
irvine, CA 92618
9492977167
MDR Report Key24601451
Report Number3019388613-2026-00088
Device Sequence Number18234230
Product Code MHX
UDI-Device Identifier00843997019163
UDI-Public(01)00843997019163(11)250327
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K223498
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 02/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9037
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/13/2026
Supplement Date Received by Manufacturer05/06/2026
Initial Report FDA Received Date03/14/2026
Supplement Report FDA Received Date05/31/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/27/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ECG ELECTRODES, (25D10, 25FBR), TWO NIBP MODULES,PSN.
Patient SexUnknown
-
-