|
(b)(6).Method: the subject mr290v autofeed humidification chamber provided as part of the rt380 adult dual heated evaqua2 breathing circuit, additional information, and photographs were requested to be provided to f&p healthcare for analysis.Results: there was no response to f&p healthcare's requests for additional information, photographs, and return of the subject device.Conclusion: based on the limited information provided by the user and without the return of the subject device, we are unable to confirm or determine the cause of the reported issue.Every mr290v vented autofeed humidification chamber is pressure tested following the manufacturing process and any holes, cracks, or leaks would be identified during this process.In addition, the pressure test is following by a visual inspection of each chamber.Any chamber which fails these tests is rejected.The subject mr290v vented autofeed humidification chamber would have met the required specifications at the time of production.The user instructions for the rt380 adult dual heated evaqua2 breathing circuit state the following: perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.Appropriate patient monitoring (e.G.Oxygen saturation) must be used at all times.Ensure appropriate ventilator or flow source alarms are set before connecting breathing sets to patient.Visually inspect breathing sets for damage (e.G.Crushed tube or cracked connector) before use and replace if damaged do not soak, wash, or sterilize this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers.Use usp sterile water for inhalation or equivalent for humidification.Do not add other substances to the water.
|