H11: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.Photo was provided for review.The investigation of the reported event is currently underway.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
H11: manufacturing review: this event is determined to be unexpected; therefore, a dhr review and manufacturing review is required.Investigation summary: the venclose generator serial number (b)(6) was received at the bd bard global service & repair.One photo of the venclose generator ((b)(4) - generator software version.Jpg) was provided by the customer.In the photo, the generator was placed the on the table.The venclose logo was displayed on the generator, and the catheter was connected on the device.Also, the home screen displaying the software version being 3.35.No other visual anomalies were noted of the internal component.Based on photo review, the reported issue is inconclusive as the exact circumstances at the time of the reported event cannot be verified from the provided photo.The venclose generator was visually inspected upon receipt and was found to be in good physical condition.The device was received with software version 3.35, and it was downgraded to 3.17 for testing.The device was functionally tested and passed the tests after downgrading the software.The short tests results are 10cm = 20.42 v 2.5cm = 6.64 v 10cm short test = pass 2.5cm short test = pass.No other anomalies were identified.Therefore, the investigation is determined to be confirmed for the redx error on software version 3.35 issue.The root cause for redx error on software version 3.35 issue could not be determined based on the provided information.Labeling review: the labelling/packaging review was not required as the reported event is not associated with a labelling, packaging, or use related issue.D4 (unique identifier (udi) #), g3, h6 (method, result, conclusion).Section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|