| Catalog Number |
MC1825 |
| Medical Device Problem Codes |
Break (1069); Failure to Obtain Sample (2533); Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records was performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.Section a through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, a patient underwent an ultrasound guided prostate biopsy procedure using maxcore instrument.During the procedure, the device was triggered, but the biopsy needle did not take a sample and sounds almost like the spring would snap/break.There was no patient reported injury.This is report 3 of 10.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, a patient underwent an ultrasound guided prostate biopsy procedure using maxcore instrument.During the procedure the device was triggered, but the biopsy needle did not take a sample and sounds almost like the spring would snap/break.There was no patient reported injury.This is report 3 of 9.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the reported detachment could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.G3, h6 (device, component, method, result, conclusion).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|