Visual inspection during the investigation, scratches on the outer housing of the valves, but no significant deformations were determined.Permeability test a permeability test has shown that both valves are permeable.During the permeability test bloody liquid were flushed out.Computer controlled test to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical position.The results show that the valves are operating within the specified tolerances in their respective relevant position.Adjustment test the valves were tested and both valves are adjustable in all pressure settings.Braking force and brake function test the brake functionality test has shown that the brake function operates as expected and the braking force is within the given tolerance.Internal inspection after dismantling of the valves, organic deposits were found in both valves.Results based on our investigation results, we can determine visible deposits in both valves.The deposits did not affect the technical properties at the time of the investigation.At the time of the examination, we were unable to determine how the aforementioned functional impairment occurred.Organic material in the patient's own cerebrospinal fluid may have temporarily affected the function and is a known side effect of hydrocephalus therapy.Not every deposit in a shunt system, depending on its location, quantity, and consistency, leads to a deterioration in performance.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|