|
(b)(4).Date sent: 3/16/2026 d4 batch #: z70225 investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis determined that the device was returned with the distal tip of the blade broken off and it was not returned device.The remaining blade portion was scratched, with evidence of contact with metal in or out of the operative field.During functional testing, an alert screen was displayed, and the device stopped activating, which is attributable to blade damage.Disassembly verified that no anomalies were found in the internal components.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples, or clips during the procedure.Once minor blade damage has occurred, subsequent activations may increase the severity of the blade damage, resulting in activation issues such as failing the pre-run test with the generator and displaying alert screens like "remove instrument from patient," ¿blade error detected,¿ or "relaxed pressure on blade," followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damaged blade can result in a broken blade.As part of the quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot/batch z70225, and no non-conformances were identified.
|