This complaint is recorded by zimmer biomet under (b)(4).The following sections were updated: b4, b5, d2, d4, d9, g3, g6, h2, h3, h4, h6 and h11.Visual examination of the returned product/provided pictures confirmed there was a plastic-like particle within the sterile packaging.The packaging does not meet the acceptance criterial per 8.400 zpc ¿ packaging materials inspection.Review of the device history record identified no deviations or anomalies during manufacturing.The root cause of the reported issue is attributed to a manufacturing issue.The event is confirmed
if any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|