| Catalog Number |
HAR9F |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an unknown procedure the device was connected to the gen11 and directly the gen indicated the tip/jaw was broken.The needed to change to a new device and then it worked.There were no patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 3/16/2026.B3: event month is unknown.D4 batch #: unknown.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 4/7/2026.D4 batch #: z7023d.Corrected data: d4 udi # and expiration date, h4.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the device was returned with no apparent damage.Functional testing was performed with the generator, and the device worked as intended; no alert screens were displayed at any time during testing.The har9f device is equipped with an eeprom that collects error codes associated with generator alert screens.During the analysis of the data collected in the eeprom while using the device, it was noted that the device triggered the "tighten assembly" alert screen.This alert occurs when the instrument is not properly assembled with the handpiece.To ensure proper assembly, the torque wrench supplied with the device should be used to tighten the blade onto the handpiece.Turn the torque wrench clockwise while holding the handpiece, and continue until it clicks twice, indicating that sufficient torque has been applied to secure the blade.For further details, please refer to the instruction for use.There were no " blade tip broke off" issues observed at any time during analysis.Due to the event description reported the investigation was confirmed.As part of the quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch z7023d, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|