MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) DURATA STS OPTIM ACTIVE FIXATION; Implantable cardioverter defibrillator (non-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) DURATA STS OPTIM ACTIVE FIXATION; Implantable cardioverter defibrillator (non-CRT) Back to Search Results
Model Number 7120/65
Medical Device Problem Codes Fracture (1260); High impedance (1291); Stretched (1601); Capturing Problem (2891); Naturally Worn (2988)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/23/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient presented in the clinic for follow up.Upon device interrogation, it was noted that the right ventricular (rv) lead exhibited high capture threshold and high pacing impedance.Chest x-ray confirmed lead fracture, stretching, and insulation was worn.Additionally, implantable cardioverter defibrillator (icd) was also migrated.No intervention was performed.There were no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION
Common Device Name
Implantable cardioverter defibrillator (non-CRT)
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key24605990
Report Number2017865-2026-05379
Device Sequence Number18734904
Product Code LWS
UDI-Device Identifier05414734502092
UDI-Public(01)05414734502092(10)2921715(17)120131
Combination Product (Y/N)Y
Initial Reporter CountryAS
PMA/510(K) Number
P950022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2012
Device Model Number7120/65
Device Lot Number2921715
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2026
Initial Report FDA Received Date03/16/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/08/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1