BIOSENSE WEBSTER INC QDOT MICRO; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
|
Back to Search Results |
|
| Catalog Number |
D139505 |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc.Or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an atrial flutter ablation procedure with a qdot micro catheter and an irrigation issue occurred.Initially, the carto 3 system was displaying high force readings when the catheter was inserted into the body.The medical team tried rezeroing several times without resolution.They then reseated the connections and rebooted the ngen¿ system and the issue resolved.They were able to complete 3 ablations but they noticed the ngen¿ pump was going 4ml even though all the settings were correct.The catheter was not irrigation properly.The catheter was replaced and the issue was resolved.The procedure was continued.No patient consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
On 18-mar-2026, the bwi product analysis lab received the device for evaluation.The product investigation was subsequently completed.It was reported that a patient underwent an atrial flutter ablation procedure with a qdot micro catheter and an irrigation issue occurred.Initially, the carto 3 system was displaying high force readings when the catheter was inserted into the body.The medical team tried rezeroing several times without resolution.They then reseated the connections and rebooted the ngen¿ system and the issue resolved.They were able to complete 3 ablations but they noticed the ngen¿ pump was going 4ml even though all the settings were correct.The catheter was not irrigation properly.The catheter was replaced and the issue was resolved.The procedure was continued.No patient consequences were reported.Device evaluation details: the device was returned to johnson & johnson medtech (j&j medtech) for evaluation.Following j&j medtech procedures, a visual inspection and irrigation test of the returned device were performed.Visual inspection revealed no damage or anomalies on the device.An irrigation test was performed, and the device was irrigating correctly, no obstructed holes were observed.No irrigation issues were observed.A manufacturing record evaluation was performed for the finished device 31579178l number, and no internal actions related to the reported complaint condition were identified.The irrigation issue reported by the customer could not be replicated during the product investigation; other issues or circumstances may have occurred during the use of the device that may have affected its performance.The instructions for use (ifu) contain the following information: purge the catheter and irrigation tubing with heparinized normal saline prior to insertion of the catheter inside the patient body.As part of johnson & johnson medtech's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
| |
|