|
It was reported that on (b)(6) 2026 at 06:49:04, the nkv-550 ventilator unexpectedly powered down.A critical red visual alarm began flashing and the ventilator ceased delivering mechanical breaths.Absence of ventilation was confirmed via waveform capnography on the patient monitor.At 06:49:28, the ventilator began to power back on.By 06:49:50, the patient was being manually ventilated by a respiratory therapist at the bedside.At approximately 06:50:00, the ventilator completed the power-on sequence.During the event, the patient became bradycardic temporarily.At approximately 06:56:00, the patient was placed on a replacement ventilator, and vital signs were reported to be stable.A review of the device logs confirmed that a momentary cpu reset occurred, with ventilation recovery occurring within 16 seconds.By design, audible and visual alarms are annunciated during a momentary system restart.
|