|
It was reported that during patient use, the ventilator rebooted and displayed the error message: ¿unit restarted, check settings, check apps activated.¿ following the reboot, the ventilator resumed ventilating the patient.As a precaution, clinical staff transferred the patient to another ventilator and removed the affected device from service.There was no patient harm reported as a result of the event.The reporting hospital declined to provide patient demographic information.A review of the device logs confirmed a momentary cpu reset, with ventilation recovery occurring within 15 seconds.By design, audible and visual alarms are annunciated when the momentary system restart occurs.
|