|
|
| Model Number |
TRIEEAXL25MT |
| Medical Device Problem Code |
Failure to Align (2522)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that preoperatively, during set-up issues and prior to firing, the anvil flipped or tilted before activation.The device was replaced with a new product.There was no patient involved.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 correction: a1, pma / 510(k) # h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection of the staple guide noted the presence of a full complement of staples.Further inspection noted that the green bar was not visible in the indicator window and the safety feature was engaged.The staple guide was observed to be attached to the instrument with no damage to the staple guide.The anvil of the instrument was observed to be not tilted and separated from the instrument.Further inspection of the anvil cutting ring showed no traces of knife impression and the ring was received intact.One of the retainer legs of the anvil was observed to be bent.Functional testing noted that due to the observed retainer leg damage of the clinical anvil, a representative anvil was attached to the center rod without difficulty prior to firing the instrument.The wing nut and safety functioned properly upon rotating the twist knob.The green bar was visible within the indicator window when the twist knob was fully closed.The instrument was applied to the appropriate test media producing acceptable results.The instrument body label was removed for evaluation of the eccentric washer assembly.Evaluation of the eccentric washer noted that it was secured.It was reported that the anvil tilted prematurely.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The evaluation detected an unreported condition: bent retainer leg.The product analysis noted evidence that the device was not used as intended.This issue may occur if the anvil is manipulated with excessive force during the attachment to the instrument, or if the anvil is mishandled during the removal of fitting accessories.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: do not grasp/clamp on the legs (open end) of the anvil/center rod assembly.Doing so could bend the legs and make it difficult to at tach/detach the anvil to the integrated trocar of the stapler.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|