| Model Number |
420093-14 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was a broken cable with a da vinci product.The customer reported event has no report of patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The instrument involved in this complaint has not been received and/or tested by failure analysis as of the date of this report.If the product is received and evaluated, a supplemental mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the da vinci product to perform failure analysis.The fenestrated bipolar forceps instrument was analyzed and found to have a derailed grip cable from its normal position at the grip input disk in the proximal housing.A visual inspection of the backend found no evidence of a cracked clamping pulley, input disk, or input shaft that would have caused the derailed cable.The probable root cause is attributed to a loss of cable tension which may have resulted from a stretched cable due to an unknown cause.
|
| |
|
Event or Problem Description
|
|
Refer to h11 for follow-up information.
|
| |
|