| Model Number |
9037 |
| Medical Device Problem Code |
No Audible Alarm (1019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the radius vsms "fails to alarm low battery." there was no patient impact or consequence reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: masimo has reached out to the customer to request the return of the device.The device has not been returned to masimo to allow an analysis to be performed.If the device is returned for evaluation or new information is obtained, a follow up report will be submitted.E1: initial reporter zip code exceeded allowable field length, initial reporter zip code is as follows: (b)(6).Health effect impact code: no adverse event; no patient impact.H3 other text: device not returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: the returned device was evaluated.The device was able to audibly and visually alarm for all enabled parameters for the duration of testing.The device issued a low battery alarm at the moment where the battery charge capacity met the specified threshold.No product performance issues related to the device's alarm system were identified when the device was functionally tested.The device was determined to be functioning as designed.E1.Initial reporter zip code exceeded allowable field length, initial reporter zip code is as follows: (b)(6).Health effect impact code: no adverse event; no patient impact.
|
| |
|
Event or Problem Description
|
|
The customer reported the radius vsms "fails to alarm low battery." there was no patient impact or consequence reported.
|
| |
|