Model Number EGIA60AXT
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/06/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D10 concomitant products: egiaustnd, egiaustnd endogia ultra univ std stap, (lot # p5k0987) egiaustnd, egiaustnd endogia ultra univ std stap, (lot # p1j0485) egiaushort, egiaushort endogia ultra univ sht stap, (lot # p4k1209) egiaushort, egiaushort endogia ultra univ sht stap, (lot # p3g0344) egia60amt, egia60amt egia 60 artic med thick sulu, (lot # n3f0595y) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, during a procedure, the four handles and two reloads failed to fire, even though the user had clicked the green button before firing.After clicking the green button, the handle was stuck and unable to move.The doctor had to switched to a different brand to complete the surgery.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6 correction: a5a, a5b h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Functional testing found that the reload was loaded into a representative instrument.The reload was partially cycled to investigate the gap between the sled and i-beam.The interlock was overridden and the reload was applied to test media.All staples were placed and test media was cleanly transected.The reload interlock was tested and found to function properly.It was reported that the instrument did not fire.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The issue of the pre-fire condition with interlock engagement may occur if the instrument firing handle had been partially compressed and released after pressing the green firing button.In this situation, the safety interlock feature will engage and prevent the reload from firing a second time by ceasing the placement of staples and tissue transection and prevent patient harm.The manufacturing records f or each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: always inspect the tissue thickness and select an appropriate staple size prior to application of the endo gia¿ universal stapler.When positioning the stapler on the application site, ensure that no obstructions, such as clips, are incorporated into the instrument jaws.Firing over an obstruction may result in incomplete cutting action and/or improperly formed staples.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer (Section G)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key24615076
Report Number1219930-2026-01679
Device Sequence Number10333825
Product Code GDW
UDI-Device Identifier10884523003253
UDI-Public(01)10884523003253(17)290930(10)N4K2168Y(20)01
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
K111825
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEGIA60AXT
Device Catalogue NumberEGIA60AXT
Device Lot NumberN4K2168Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2026
Supplement Date Received by Manufacturer06/11/2026
Initial Report FDA Received Date03/17/2026
Supplement Report FDA Received Date06/15/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/29/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Patient Age28 YR
Patient SexMale
-
-