• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMVIE US CORP LLC IMP,TSV,4.1MM,SBM,10; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMVIE US CORP LLC IMP,TSV,4.1MM,SBM,10; DENTAL IMPLANT Back to Search Results
Catalog Number TSV4B10
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/31/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A1: patient identifier unknown / not provided.A2: age at time of the event unknown / not provided.A3: gender unknown / not provided.A4: patient weight unknown / not provided.D9: device availability unknown / not provided.
 
Event or Problem Description
Doctor reported that during implant placement of implant 4.1 x10, the mount did not detach from the implant at tooth site # 13.The implant was removed and was replaced with another one of the same size.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4) the following fields have been updated: b4: date of this report b5: describe event or problem g3: date received by manufacturer g6: type of report h1: type of reportable event h2: follow up type h3: device evaluated by manufacturer h6: adverse event problem h11: additional narrative zimvie did not receive one (1) tsv4b10, (imp,tsv,4.1mm,sbm,10) for evaluation.Visual inspection could not be performed.The investigation has been performed based on the available information using the item number, device history record (dhr) review, and risk management file (rmf).Functional testing to recreate the reported event could not be performed due to the nature of the device & event.This complaint refers to the specific device being investigated for this complaint record.Dhr review was completed for the subject lot number 1286041.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 1286041 for similar events and no other complaint was identified.Review completed utilizing keywords: ¿does not disengage/release¿ based on the investigation and risk management file review, the most likely root causes determined from the investigation are incorrect techniques used during implant placement - customer error.Therefore, based on the available information, a device malfunction could not be verified.Without device receipt, the reported event is non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMP,TSV,4.1MM,SBM,10
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
mark mashburn
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key24615120
Report Number0001038806-2026-01429
Device Sequence Number9586123
Product Code DZE
UDI-Device Identifier00889024019225
UDI-Public(01)00889024019225(17)290911(10)1286041
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K072589
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 04/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberTSV4B10
Device Lot Number1286041
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/17/2026
Supplement Report FDA Received Date04/20/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/12/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-