Model Number 453977
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/02/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Combination product: yes.Neither the affected device nor the angiographic material was returned.Therefore, no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the device was manufactured according to specifications and passed all in-process and final inspections.Based on the conducted investigations, no material or manufacturing related root cause could be determined.
 
Event or Problem Description
The orsiro mission drug-eluting stent system was selected for treatment of a severely calcified lesion in a severely tortuous segment of the proximal lcx.It is unknown if a pre-dilation of the lesion was performed.The affected device was introduced into the patient's body but could not cross the lesion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO MISSION US, 3.5/40
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG
ackerstrasse 6
bulach, zurich 8180
SZ  8180
Manufacturer (Section G)
BIOTRONIK AG
ackerstrasse 6
bulach, zurich 8180
SZ   8180
Manufacturer Contact
jana burger
ackerstrasse 6
bulach, zurich 8180
SZ   8180
MDR Report Key24615210
Report Number8043892-2026-03032
Device Sequence Number9834174
Product Code NIQ
UDI-Device Identifier07640130456334
UDI-Public(01)07640130456334(17)280825(10)07250685
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number453977
Device Lot Number07250685
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/03/2026
Initial Report FDA Received Date03/17/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/27/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-