MAUDE Adverse Event Report: CLEARSTREAM TECHNOLOGIES LTD. HALO ONE THIN-WALLED GUIDING SHEATH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CLEARSTREAM TECHNOLOGIES LTD. HALO ONE THIN-WALLED GUIDING SHEATH Back to Search Results
Catalog Number HLO62535H
Medical Device Problem Codes Difficult to Insert (1316); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/24/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: as the lot number for the device was provided, a review of the device history records was performed.The return of the sample is pending.The investigation of the reported event is currently underway.Section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
On (b)(6) 2026, a patient underwent a recanalization procedure utilizing the halo one device.During the procedure a crack was observed at the tip of the dilator, resulting in increased difficulty during insertion.The procedure was subsequently completed using an alternative device.There was no reported patient injury.
 
Event or Problem Description
On (b)(6) 2026, a patient underwent a recanalization procedure utilizing the halo one device.During the procedure a crack was observed at the tip of the dilator, resulting in increased difficulty during insertion.The procedure was subsequently completed using an alternative device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H11: manufacturing review: a complaint history review was performed.This is the fourth complaint reported for this product/lot number combination.However, the device history records were reviewed and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the halo device was returned for evaluation.Two splits were present on the dilator commencing at the tip and measured 7mm and 4mm.The sheath was undamaged.Therefore, the result of the investigation is confirmed for the reported dilator split and insertion issues.The root cause for the reported dilator split and insertion issues could not be determined based upon the available information received from the field communications and device evaluation.Labeling review: the instruction for use for this lifestream device was reviewed and the following sections are applicable indication for use the halo one thin walled guiding sheath is indicated for use in peripheral arterial and venous procedures requiring percutaneous introduction of intravascular devices the halo one thin walled guiding sheath is not indicated for use in the neurovasculature or the coronary vasculature contraindications there are no known contraindications for the halo one thin walled guiding sheath warnings 1 contents supplied sterile using ethylene oxide eto non pyrogenic do not re use reprocess or resterilize this device is intended for single use only 2 do not resterilize after resterilization the sterility of the product is not guaranteed because of an indeterminable degree of potential pyrogenic or microbial contamination which may lead to infectious complications cleaning reprocessing andor resterilization of the present medical device increases the probability that the device will malfunction due to potential adverse effects on components that are influenced by thermal andor mechanical changes 4 visually inspect the packaging to verify that the sterile barrier is intact do not use if the sterile barrier is opened or damaged 5 use the sheath prior to the use by date specified on the package 6 do not advance the guidewire sheathdilator procedural device or any component if resistance is met without first determining the cause and taking remedial action 8 the halo one thin walled guiding sheath has not been evaluated for use in the neurovasculature or the coronary vasculature 10 only advance or retract the sheath with the dilator inserted and only advance or retract the sheath and dilator while placed over a properly sized guidewire precautions 1 the halo one thin walled guiding sheath shall only be used by trained physicians access procedures should be conducted under fluoroscopic guidance with appropriate xray equipment and or ultrasound 4 the minimum acceptable sheath french size is printed on the package label do not attempt to pass devices through a smaller size sheath introducer than indicated on the device label 6 carefully inspect the sheath prior to use to verify that the sheath has not been damaged during shipment and that its size shape and condition are suitable for the procedure for which it is to be used do not use if product damage is evident 8 insert dilator into the center of the sheath valve forced insertion of the dilator which misses the center of the valve may cause damage and result in blood leakage 9 advance or withdraw the sheath slowly if resistance is met do not advance or withdraw until the cause of resistance is determined 12 when using procedural devices close to the tip of the sheath care must be taken to ensure the active mechanism portion of the procedural device eg balloon stent zone material removal section of atherectomy device is not within the tip of the sheath the radiopaque marker is located within 5 mm of the end of the tip but does not mark the true distal tip of the sheath 14 ensure to deactivate the procedural device prior to removal through the sheath 15 take care when back loading the tip of the dilator over the guidewire to avoid damage to the dilator 16 ensure the dilator is securely connected with the sheath prior to advancing otherwise only the sheath may advance into the vessel and the sheath tip may cause damage to the vessel 17 ensure the dilator is in place within the sheath before advancing the sheath as otherwise damage may be caused to the vessel 18 do not place sutures on the sheath tubing since this may restrict accessflow through the sheath when puncturing suturing or incising near the sheath be careful not to damage the sheath proper functioning of the sheath depends on its integrity care should be used when handling the sheath 20 if resistance is felt during postprocedure withdrawal of the procedural device it is recommended to remove the procedural device guidewire and sheath as a single unit 21 in order to activate the hydrophilic coating it is recommended to wet the halo one thin walled guiding sheath with heparinized saline solution immediately prior to its insertion in the body failure to activate the coating may lead to sub optimal trackability of the sheath to maintain lubricity this surface must be kept completely wet 22 proper functioning of the halo one thin walled sheath depends on its integrity care should be used when handling the sheath damage may result from kinking stretching or forceful wiping of the halo one thin walled guiding sheath do not continue to use the sheath if the shaft has been bent or kinked storage store in a cool dry dark place rotate inventory so that the catheter and other dated products are used prior to the use by date do not use if packaging is damaged or opened directions for use halo one thin walled guiding sheath preparation 1 verify the french size is suitable for the procedure and can accommodate the required procedural devices as labeled figure 3 remove sheath and dilator from package 2 prior to use the air in the sheath and dilator should be removed to facilitate purging the device is packaged with the dilator inside the sheath in a reverse direction allowing both to be flushed simultaneously select a syringe or inflation device with a 10 ml or larger capacity and fill approximately half of it with sterile saline solution prepare the lumen by connecting the syringe or inflation device containing the solution into the stopcock on the side port of the sheath and flush with the sterile heparinized saline solution figure 4 close the stopcock to maintain the air tightness following flushing 3 prior to use the reverse loaded dilator must be removed from the distal end of the sheath if not already completed at step 2 the air in the dilator lumen should be removed to facilitate purging select a syringe or inflation device with a 10 ml or larger capacity and fill approximately half of it with sterile saline solution prepare the lumen by connecting the syringe or inflation device containing the solution into the luer connector of the dilator hub and flushing with the sterile heparinized saline solution figure 7 4 insert the provided vessel dilator through the hemostatic valve and click the dilator hub into place in the valve housing figures 5 and 6 5 in order to activate the hydrophilic coating where labeled it is recommended to wet the halo one thin walled guiding sheath with sterile saline solution immediately prior to its insertion in the body use of the halo one thin walled guiding sheath 6 identify the insertion site including but not limited to radial femoral popliteal tibial or pedal access sites and prepare the site using proper aseptic technique and local anesthesia as required 7 at the operators discretion make a small skin incision at the puncture site with a surgical scalpel recommended for scar tissue at the access site 8 backload the distal tip of the halo one thin walled guiding sheath dilator over the prepositioned guidewire and advance the tip to the introduction site 9 advance the dilator and the sheath through the skin and into the vessel grasp the sheath and dilator assembly close to the skin while advancing to avoid buckling employ a rotating motion while advancing as required note if using a hydrophilic guidewire ensure that it is kept hydrated with sterile heparinized saline at all times 10 carefully advance the dilator and the sheath over the wire to the site required within vessel the radiopaque marker identifies the sheath tip location under fluoroscopy 11 disconnect the dilator hub from the valve by bending to the side to unsnap from the valve cap figure 8 remove the dilator slowly while holding the sheath in place and ensuring the guidewire remains in place figure 9 as required 12 load the procedural device over the prepositioned guidewire 13 advance the procedural device carefully into the centre of the valve diaphragm and through the sheath to the treatment site figure 10 while maintaining the sheath position 14 position the procedural device relative to the lesion to be treated figure 11 ensuring the active mechanism portion of the procedural device is not within the sheath 15 following use ensure the procedural device is fully deactivated before carefully withdrawing the procedural device through the sheath while maintaining the sheath position 16 after the procedure is completed insert the dilator over the wire into the sheath 17 slowly withdraw the sheath and dilator as a unit and then remove the guide wire g3, h6 (component, method, result, conclusion).Section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HALO ONE THIN-WALLED GUIDING SHEATH
Common Device Name
GUIDING SHEATH
Manufacturer (Section D)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford N A
EI  N A
Manufacturer (Section G)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford N A
EI   N A
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key24615308
Report Number9616666-2026-00045
Device Sequence Number21377802
Product Code DYB
UDI-Device Identifier00801741184871
UDI-Public(01)00801741184871(11)250123(17)271231(10)CMKN0147
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K192313
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberHLO62535H
Device Lot NumberCMKN0147
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/24/2026
Supplement Date Received by Manufacturer03/31/2026
Initial Report FDA Received Date03/17/2026
Supplement Report FDA Received Date04/07/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/23/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-