| Model Number |
CS1 |
| Medical Device Problem Code |
No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/16/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Ge healthcare¿s investigation into the reported occurrence is ongoing.A follow-up report will be issued when the investigation has been completed.Block a: no report of patient involvement.Legal manufacturer: hcs research park - 9900 innovation drive usa wauwatosa, wi 53226.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a malfunction resulting in oxygen failure during service.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The supply valve was replaced to resolve the issue.
|
| |
|
Search Alerts/Recalls
|