• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. VCL+ UD 36IN 1 S/A CT-1; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VCL+ UD 36IN 1 S/A CT-1; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number VCP947H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an unknown procedure on an unknown date and suture was used.During a orthopedic procedure the needles are popping off.No patient consequences.There were no patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 3/17/2026 h6 component code: g07002 - pending evaluation of returned device this report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, the following was obtained: * how many devices demonstrated the alleged deficiency during this procedure? they didn¿t give an exact number, but it was enough for a noticeable amount to call me and stop using the lot number.* when did the alleged deficiency occur (removal from package / during handling prior to use on patient/ during passage through tissue / during tying / post-op)? if different, please explain.Passing through the tissue * please provide the status of the device(s) as it has not been received for analysis.If the device has been shipped, please provide the shipment tracking details.Device has been shipped, will locate tracking information.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 4/14/2026.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device problem found.Additional information: d9, h3, h6.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: the product was returned to ethicon for evaluation.One open box with twenty-nine (29) representative unopened samples and five empty open foils were received for analysis.Product code vcp947h.The complaint sample was not received for evaluation.Upon initial inspection of the samples, no external damages were evident on the exterior packaging.Following the sampling plan, a visual inspection was conducted on thirteen (13) samples.The packets were opened, and swage and attachment areas appeared as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related no defects were observed during evaluation.A functional test was performed using instron equipment and the pull force result was above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL+ UD 36IN 1 S/A CT-1
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key24616342
Report Number2210968-2026-02674
Device Sequence Number21226284
Product Code GAM
UDI-Device Identifier10705031056169
UDI-Public(01)10705031056169(17)300731(10)10A3K5
Combination Product (Y/N)Y
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K032420
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberVCP947H
Device Lot Number10A3K5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/18/2026
Supplement Date Received by Manufacturer03/30/2026
Initial Report FDA Received Date03/17/2026
Supplement Report FDA Received Date04/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/02/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-