Catalog Number T3ST3210
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Osteolysis (2377); Swelling/ Edema (4577)
Date of Event 12/16/2025
Type of Reportable Event Serious Injury
Event or Problem Description
The doctor reports that the dental implant located in dental position number 25 failed due to peri-implantitis.The doctor reports in the per that the procedure was not concluded by placing another implant.Pain, edema and inflammation were reported as a consequence of the event.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A4: patient weight: unknown / not provided.E1: reporter name: unknown / not provided.H6: event problem codes: patient code: 1994, 1932.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T3® PRO NON-PLATFORM SWITCHED TAPERED IMPLANT 3.25MM (D) X 10MM (L)
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
mark mashburn
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key24616861
Report Number0001038806-2026-01437
Device Sequence Number13155646
Product Code DZE
UDI-Device Identifier00844868046868
UDI-Public(01)00844868046868(17)281021(10)2120004889
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
K213672
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 03/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberT3ST3210
Device Lot Number2120004889
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date03/17/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/23/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ILPC342U.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient SexMale
Patient EthnicityNon Hispanic
-
-