It was learned through implant patient registry and records that a patient with a 25mm 3300tfx aortic valve was explanted after an implant duration of 8 years, 8 months due to severe stenosis.Patient presented with heart failure.The explanted valve was replaced with a 23mm 11500a inspiris resilia aortic valve.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional 'customer complaint'.The information reported may or may not be related to the edwards device.
|
|
H11: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
|
H11: additional manufacturer narrative: updated sections : b7, g3, g6, h2,h6 ( type of investigation) corrected sections : b5, h6 ( component code , device code, investigation findings and investigation conclusions).Svd is a logical and expected consequence of the chemical, mechanical, and immunological processes associated with bhv implantation.The common endpoint of all these factors is calcification and degradation.Svd is an acquired condition, intrinsic to bioprosthetic valves, characterized by deterioration of the leaflets or supporting structures, leading to thickening, calcification, tearing, or disruption of the prosthetic valve materials.These changes result in valvular dysfunction manifesting as stenosis and/or regurgitation with a consequent drop in hemodynamic efficacy of the valve.Despite the advancements in the heart valve industry, svd commonly begins 7- 8 years after implantation, with a considerably increased rate in patients with pertinent comorbidities and/or an implant duration of greater than 10 years.Prior investigations and extensive literature review have shown that svd is predominantly related to patient factors and implant duration rather than to manufacturing issues.Events related to device design, manufacturing, or use error tend to manifest at an earlier implant duration.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.Based on the information available, the most likely root cause is patient factors, including an implant duration of 7 or more years, hyperlipidemia (hld), chronic kidney disease (ckd), diabetes mellitus (dm), and coronary artery disease (cad).The device history record (dhr) review was completed, and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
|
It was learned through implant patient registry and records that a patient with a 25mm 3300tfx aortic valve was explanted after an implant duration of 8 years, 8 months due to degeneration with thickened leaflets and severe stenosis.Patient presented in heart failure with chest pain.The explanted valve was replaced with a 23mm 11500a inspiris resilia aortic valve.Intraoperatively, previous aortic valve showed clear evidence of degeneration.Valve was explanted and replaced with a 23mm 11500a aortic valve.The patient also underwent mvr of native valve.Tee demonstrated good seating with no pvl.Patient was sent to opacu in stable but critical condition and discharged on pod#19.
|