• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH INSER DEV F/ZERO-P VA; INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, CERVICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH INSER DEV F/ZERO-P VA; INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, CERVICAL Back to Search Results
Catalog Number 03.647.963
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/05/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported during preparation, it was found that the locknut does not rotate correctly.The thread appears to be defective.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the product was returned to depuy synthes for evaluation.Visual inspection of the returned device found normal use wear.Threads of the knurl-nut were stripped and during functional test a small resistance was noted due to this condition.Additionally, the distal end of the handles was deformed.The observed condition of the device was consistent with a component failure that was caused by exposure to unintended forces.Properly handling and attention to the approved use of the device diminishes the risk of failure.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the inser dev f/zero-p va would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Dimensional inspection: n/a.Device history: part number: 03.647.963.Lot number: 31115p8 (wo# (b)(4)).Product code: n/a.Date of manufacture: 27-sep-2024.Manufacturing location: tuttlingen.Part expiration date: n/a.Nonconformance noted: n/a.Dhr record review: a manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSER DEV F/ZERO-P VA
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, CERVICAL
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
TUTTLINGEN
unter hasslen 5
tuttlingen 78532
GM   78532
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
9083863534
MDR Report Key24617615
Report Number8030965-2026-02566
Device Sequence Number15863178
Product Code OVE
UDI-Device Identifier07611819404458
UDI-Public(01)07611819404458(10)31115P8
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K112068
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.647.963
Device Lot Number31115P8
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/11/2026
Initial Report FDA Received Date03/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/27/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-