| Model Number |
G5A-80A |
| Medical Device Problem Code |
Failure to Charge (1085)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device prompted a "hv charger test capacitor not holding charge" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Review of the device activity logs makes it clear the device presented an error where the end result was that the device code indicator was red and was beeping.The device will communicate to the user that it is not rescue ready by providing an auditory beep every 30 seconds and displaying a red rescue ready (nvi) indicator.We can identify from the log that this device appears to have been left idle in a non-usable state since may of 2025.Zoll medical corporation evaluated the device and the customer's report was not replicated or confirmed.The device was put through extensive testing including daily/weekly/monthly testing, on/off current testing, patient/circuit impedance testing and shock testing without duplicating the report.The main board was replaced as a precaution.The device was recertified and returned to the customer.No trend is associated with reports of this type.
|
| |
|
Search Alerts/Recalls
|