| Catalog Number |
PSE45A |
| Medical Device Problem Codes |
Failure to Form Staple (2579); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a pulmonary procedure, it was observed that there was bleeding after a fire of white reload (not the 1st fire).The volume of bleeding was about 300cc.Suture sewing was used to control the bleeding.After a fire of a blue reload, noted malformed staples.Changed to another competitive device to complete the procedure.There was no patient consequence nor surgical delay reported.No additional information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 3/17/2026.D4: batch #unk.Additional information was requested, and the following was obtained: how was the bleeding controlled? suture sewing.How much blood was lost (ml)? less than 300 cc.Did the patient require a transfusion? no.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device pse45a lot number a98m70 and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 4/8/2026.D4: batch #709d34.Corrected data: d1.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device visual analysis of the returned sample determined that one pse45a device was returned with no apparent damage and with one gst45w and one gst45b reloads present.The reloads were received fully fired.The device was tested for functionality in the straight position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples met the staple release criteria.The event described could not be confirmed as the device performed without any difficulties noted.No malformed staples were observed.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 709d34, and no non-conformances were identified.
|
| |
|