• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN SURGICAL V-LOC 180; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN SURGICAL V-LOC 180; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC Back to Search Results
Model Number VLOCL0316
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/10/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D10.Concomitant product: vlocl0326, vlocl0326 v-loc* 180 0 grn 45cm gs21x12, (lot number: a4l1614vy).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, during open knee replacement, at the time of joint capsule suturing, upon opening the package, two sutures were found with detached needle.The suture was replaced to resolve the issue.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted one suture and one needle.The suture was returned fully wound within the retainer.The needle was returned disengaged from the suture.Upon microscopic inspection of the needle, normal crimp and swage marks were noted.No suture material was observed within the swage.The foil pouch was inspected and no signs of damage or abnormalities were identified.It was reported that the needle was detached.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-LOC 180
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC
Manufacturer (Section D)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer (Section G)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key24620721
Report Number1219930-2026-01694
Device Sequence Number10137730
Product Code GAM
UDI-Device Identifier10884521131316
UDI-Public(01)10884521131316(17)270930(10)A4K1882VY(20)01
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K091087
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVLOCL0316
Device Catalogue NumberVLOCL0316
Device Lot NumberA4K1882VY
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/12/2026
Supplement Date Received by Manufacturer06/01/2026
Initial Report FDA Received Date03/17/2026
Supplement Report FDA Received Date06/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/16/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Patient SexMale
-
-