BOSTON SCIENTIFIC CORPORATION MUSTANG?; STENTS, DRAINS AND DILATORS FOR THE BILIARY DUCTS Back to Search Results
Model Number H74939171052070
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/20/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 70% stenosed target lesion was located in the moderately tortuous and mildly calcified radial vein.A 5.0 x 200, 75cm mustang balloon was advanced for dilation.However, during the first inflation at 10 atmospheres for 30 seconds, the middle position of the balloon ruptured.The device was removed without any problem using the normal method, and the procedure was completed with another of the same device.There were no patient complications as a result of this event, and the patient was in good condition post-procedure.
 
Additional Manufacturer Narrative
D2b - pro code (product code): fge, lit g4 - premarket / 510(k) #: k103751, k110122, k141521.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG?
Common Device Name
STENTS, DRAINS AND DILATORS FOR THE BILIARY DUCTS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
mark mischke
4100 hamline ave n
arden hills, MN 55112
7634941632
MDR Report Key24621438
Report Number2124215-2026-14757
Device Sequence Number18698100
Product Code FGE
UDI-Device Identifier08714729794202
UDI-Public(01)08714729794202(17)280729(10)37087882
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K103751
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH74939171052070
Device Catalogue NumberH74939171052070
Device Lot Number0037087882
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2026
Initial Report FDA Received Date03/17/2026
Date Device Manufactured07/30/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-