• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD ACETLR CUP HAP 54MM W/ IMPTR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD ACETLR CUP HAP 54MM W/ IMPTR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74120154
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Cellulitis (1768)
Date of Event 10/10/2008
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that after a left bhr performed on (b)(6) 2008, patient was hospitalized due to cellulitis around the perioperative site on (b)(6) 2008.This adverse event was treated with keflex.No further information available.
 
Additional Manufacturer Narrative
Internal reference number: case-(b)(4).H3, h6: this complaint was opened by smith+nephew to document a patient complication reported that includes reference to the use of a smith+nephew product without evidence about a specific product problem.The reported complication relates to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).Additional information: d10: concomitant medical products and therapy dates, h6: investigation conclusions and type of investigation.H11:as of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.A review of the historical complaints data for the alleged devices was performed using batch numbers, part numbers and the reported failure modes to evaluate patterns of repeated failures or defects.Similar complaints have been identified for cup.This failure will continue to be monitored via routine trending.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All released devices involved met manufacturing specifications upon release for distribution.The review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, no prior applicable escalation actions were identified.The available medical documents were reviewed.The adverse event was likely procedure related and the device had no influence on the event.Cellulitis is a risk of any surgical procedure due to impaired skin integrity and/or weakened immune system.The root cause can likely be ascribed to non-device related factors.Based on the limited information provided we are unable to speculate on specific factors known to contribute to the alleged fault.Should the devices or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACETLR CUP HAP 54MM W/ IMPTR
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key24624250
Report Number3005975929-2026-00006
Device Sequence Number8688672
Product Code NXT
UDI-Device Identifier03596010502582
UDI-Public03596010502582
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P040033
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number74120154
Device Lot Number090263
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/19/2026
Supplement Date Received by Manufacturer04/20/2026
Initial Report FDA Received Date03/17/2026
Supplement Report FDA Received Date04/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
- KEFLEX.; PART#:74121146, LOT#:085308.
Outcome Attributed to Adverse Event Other; Hospitalization;
Patient Age43 YR
Patient SexMale
-
-