Tw # (b)(4).Updated sections: b4, d4-udi#, g3, g6, h2, h3, h6, h11.The device was returned to the factory for evaluation on 03/16/2026.An investigation was conducted on 03/17/2026.A visual inspection was conducted.Signs of clinical use and no evidence of blood was observed.There were no visual defects observed on the intact cable.Both connector ends were observed to be intact.The cable was able to be connected to the adaptor with no physical or visual difficulties.The reference harvesting device was attached to the cable with no physical or visual difficulties observed.An electrical evaluation was conducted.A pre-cautery test was performed per the instruction for use (ifu) with a reference hemopro 2 evh tool, reference adapter and reference power supply set at level 3.0.The device passed the pre-cautery test.The power supply did emit an audible beeping sound and the evh reference tool did produce steam or heat.An activation and transection capability test was performed over four (4) repetitions using "max life test method stm2048073.The device successfully transected tissue four (4) times.Based on the returned condition of the device as well as the evaluation results, the reported failure "failure to deliver energy" was not confirmed.The reported device is an oem device, therefore, a lot history review was not applicable.A lot number was not provided and the specific product lot number cannot be identified from a ship history, therefore it is impossible to obtain a certificate of conformance.
|