• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC HEMOPRO2 EXTENSION CABLE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC HEMOPRO2 EXTENSION CABLE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VH-4030
Medical Device Problem Code Failure to Deliver Energy (1211)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/25/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
 
Event or Problem Description
The hospital reported that the hemopro 2 extension cable is not working.The power supply power setting is unknown.No power energy to disposable.Changing out the power supply was not mentioned/unknown.Patient in operating room no patient effects or delays reported.Opened up another hp2 extension cable to complete case.
 
Additional Manufacturer Narrative
Tw # (b)(4).Updated sections: b4, d4-udi#, g3, g6, h2, h3, h6, h11.The device was returned to the factory for evaluation on 03/16/2026.An investigation was conducted on 03/17/2026.A visual inspection was conducted.Signs of clinical use and no evidence of blood was observed.There were no visual defects observed on the intact cable.Both connector ends were observed to be intact.The cable was able to be connected to the adaptor with no physical or visual difficulties.The reference harvesting device was attached to the cable with no physical or visual difficulties observed.An electrical evaluation was conducted.A pre-cautery test was performed per the instruction for use (ifu) with a reference hemopro 2 evh tool, reference adapter and reference power supply set at level 3.0.The device passed the pre-cautery test.The power supply did emit an audible beeping sound and the evh reference tool did produce steam or heat.An activation and transection capability test was performed over four (4) repetitions using "max life test method stm2048073.The device successfully transected tissue four (4) times.Based on the returned condition of the device as well as the evaluation results, the reported failure "failure to deliver energy" was not confirmed.The reported device is an oem device, therefore, a lot history review was not applicable.A lot number was not provided and the specific product lot number cannot be identified from a ship history, therefore it is impossible to obtain a certificate of conformance.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOPRO2 EXTENSION CABLE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 07470
MDR Report Key24625476
Report Number2242352-2026-0000189
Device Sequence Number9068616
Product Code GEI
UDI-Device Identifier00607567700925
UDI-Public(01)00607567700925
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVH-4030
Device Catalogue NumberVH-4030
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/25/2026
Supplement Date Received by Manufacturer03/30/2026
Initial Report FDA Received Date03/18/2026
Supplement Report FDA Received Date03/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NO OTHER CONCOMITANT PRODUCTS WERE NEEDED.
Patient SexUnknown
-
-