Model Number PXB35-05-17-080
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 02/19/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that during pre-renal stenting for laser fenestrated endovascular aortic repair in a patient, multiple issues occurred involving the stent visi pro 035.The procedure targeted the proximal right and left renal arteries, which exhibited minimal tortuosity and calcification, with a vessel diameter of 4.8mm and no lesion stenosis.The first stent dislodged during preparation prior to insertion into the patient.A second stent dislodged during use in the patient but was removed successfully in tandem with the delivery system without injury or intervention.A third stent also dislodged during use in the patient, and a snare attempt was made to remove the dislodged stent.In each instance, a new device was used to complete the procedure.No patient symptoms, complications, adverse outcomes, illnesses, infections, or deaths were reported.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Additional information: the second stent was removed successfully from the patient.The third stent is retained in the patient.After trying to snare it, it embolized to the left internal iliac artery and remains.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISI-PRO BALLOON-EXPANDABLE STENT SYSTEM
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key24626289
Report Number2183870-2026-00114
Device Sequence Number12910506
Product Code NIO
UDI-Device Identifier00643169787575
UDI-Public(01)00643169787575(17)270703(10)B755274
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P030045
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPXB35-05-17-080
Device Catalogue NumberPXB35-05-17-080
Device Lot NumberB755274
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/19/2026
Supplement Date Received by Manufacturer03/18/2026
Initial Report FDA Received Date03/18/2026
Supplement Report FDA Received Date04/02/2026
Date Device Manufactured07/03/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-