| Model Number |
GIA8038L |
| Medical Device Problem Code |
Sharp Edges (4013)
|
| Health Effect - Clinical Code |
Unspecified Tissue Injury (4559)
|
| Date of Event |
02/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that intraoperatively, when loading the next cartridge, the staples actually came up and punctured the nurse's skin.Then after successfully firing two reloads, some staples were protruding from the cartridge and pricked the nurse's hand.No intervention was done.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: gia8038s, gia8038s gia 80-3.8sgl use reload stap (lot#p5f1143); gia8038l, gia8038l gia 80-3.8sgl use loadingunit (lot#p5e1179) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.For functional te sting, the firing knob advanced and retracted without difficulty.The knife cut completely and cleanly.The remaining staples deployed and formed properly.It was reported that intraoperatively, when loading the next cartridge, the staples actually came up and pun ctured the nurse's skin.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur if the firing stroke is not complete and the firing knob is not fully retracted after firing or if the instrument is applied against an excessive amount of tissue during the clinical application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: the instrument should not be used on tissue such as liver or spleen where compressibility is such that closure of the instrument would be destructive.Tissue thickness should be carefully evaluated before firing any stapler.If the tissue can not comfortably compress to the specified minimum requirement, or compresses to less than this requirement, the tissue may be too thick or too thin for the selected staple size.Do not rotate the firing knob during firing.Rotating the knob during firing may cause damage and possible instrument malfunction.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|