• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER LEIBINGER FREIBURG SCR/2.3X8MM BONE CROSS PIN; IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER LEIBINGER FREIBURG SCR/2.3X8MM BONE CROSS PIN; IMPLANT Back to Search Results
Catalog Number 50-23408
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 01/01/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported infection could not be confirmed due to missing information; investigation of returned screws showed damage from high mechanical forces and instrument handling, with no device or manufacturing issues identified.Infection is a known ifu listed adverse effect.Based on the investigation, there is no indication for an incorrectly working product or any design, material, or manufacturing related issue.Therefore, no corrective and/or preventive actions are deemed necessary at this time.
 
Event or Problem Description
It was reported that a patient developed osteomyelitis with a non-union fracture and a screw was broken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCR/2.3X8MM BONE CROSS PIN
Common Device Name
IMPLANT
Manufacturer (Section D)
STRYKER LEIBINGER FREIBURG
boetzingerstr. 41
freiburg D-791 11
GM  D-79111
Manufacturer (Section G)
STRYKER LEIBINGER FREIBURG
boetzingerstr. 41
freiburg D-791 11
GM   D-79111
Manufacturer Contact
elena drewenskus
boetzingerstr. 41
freiburg D-791-11
GM   D-79111
7614888208
MDR Report Key24626618
Report Number0008010177-2026-00035
Device Sequence Number8985793
Product Code JEY
UDI-Device Identifier37613154184232
UDI-Public37613154184232
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Catalogue Number50-23408
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/17/2026
Initial Report FDA Received Date03/18/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-