| Catalog Number |
72200616F |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Overheating of Device (1437); Mechanical Jam (2983); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that during a hip arthroscopy procedure, the mdu was generating noise and became hot.The procedure was resumed, without any delay, using an equivalent s+n back-up.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: case-(b)(4).H2: additional information on h6 (medical device problem code).H3, h6: the reported device was received for evaluation.A visual inspection was performed on the exterior of the device, and no physical damage was observed.A functional evaluation was performed on the returned device and found the device grew warm and a blade stall error message was found.It was also noted device ma was above 960.Further investigation revealed stiffness in the drive fork when rotated.These findings are related to the reported event.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with a component failure.Factors that could have contributed to the failure include a blade stall condition that will result in increased current draw from the control unit which will heat the motor and hand piece housing.This can be the result of gearbox corrosion caused by cleaning and sterilization methods and the chemicals involved.Based on this investigation, the need for corrective action is not indicated.
|
| |
|