• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VIANT MEDICAL, LLC VIANT MEDICAL, LLC; EZ CLEAN¿ REAMER HANDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VIANT MEDICAL, LLC VIANT MEDICAL, LLC; EZ CLEAN¿ REAMER HANDLE Back to Search Results
Model Number T2756
Medical Device Problem Codes Mechanical Problem (1384); Unstable (1667); Material Deformation (2976); Material Integrity Problem (2978); Material Twisted/Bent (2981); Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/16/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: the customer has indicated the complaint sample will be returned to viant for evaluation but has not yet been received to date.Once the complaint sample is received, it will be investigated and a follow-up medwatch 3500a emdr will be submitted.G2: complaint information provided by distributor, (b)(4).Complaint source is foreign as event occurred in canada.
 
Event or Problem Description
It was reported during an unknown patient procedure that the handle did not ream straight as if it was curved a little.No consequences or impact to patient.
 
Additional Manufacturer Narrative
H3: the complaint sample was returned incomplete to viant for evaluation and the reported event is confirmed.The straight reamer handle power adaptor is slightly bent leading to vibration during rotation.The teflon sleeve was not returned with the assembly rendering the device incomplete and unable to assemble as intended.The stryker/zimmer-hall power adaptor end was observed to be slightly bent out of place.A bent/deformed power adaptor causes a vibration/non-linear (eccentric) reaming motion of the device when connected and powered by the surgical drill.Bending of this nature would not occur during normal intended use and likely would had occurred from applying non-linear load (off-axis force applied).Thus, the complaint sample had been misused.It is unknown what adaptor or power source (neither provided by viant) was connected to the device.The current ifu sent with this device today, man-004006 rev.A, states the following: end of life is determined by wear and damage due to intended use, visually inspect for damage and wear.If the instrument is damaged and worn it is considered at the end of its life and should be discarded, where instruments form part of a larger assembly, check assembly with mating components, check hinged instruments for smooth movement, viant devices should only be used by qualified personnel fully trained in the use of the surgical instruments and the relevant surgical procedures, do not modify viant instruments in any way and handle with care at all times.Surface scratches can increase wear and the risk of corrosion, manual surgical instruments have a limited life-span which is determined by wear or damage due to repeated intended use.When a surgical instrument reaches the end of its functional life, clean the instrument of any and all biomaterial/biohazards and safely discard the instrument in accordance with applicable laws and regulations.The device history records (dhr) were reviewed and found no related manufacturing deviations or anomalies that would have contributed to the reported event.The device had experienced approximately 1.57 years of use.It is unknown as to how many surgical procedures (cycles) this device had experienced throughout its life in the field.The viant risk management files were reviewed and the failure mode was identified and mitigated to the lowest possible level.A trend analysis was performed and the failure rates fall within the occurrence rate identified in the viant risk management files.In conclusion, the reported event is confirmed as the returned straight reamer handle power adaptor is slightly bent leading to vibration during rotation.From the investigation performed, the root cause is attributed to user error (misuse) as bending of this nature would not occur during normal intended use and likely would had occurred from applying non-linear load (off-axis force applied).No further investigation with regard to this complaint is required at this time.Viant will continue to monitor for trends.D9, g3, h2: added date device returned to manufacturer.H6: updated component, type of investigation, investigation findings, & conclusions codes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIANT MEDICAL, LLC
Common Device Name
EZ CLEAN¿ REAMER HANDLE
Manufacturer (Section D)
VIANT MEDICAL, LLC
4545 kroemer road
fort wayne IN 46818
Manufacturer (Section G)
VIANT MEDICAL, LLC
4545 kroemer road
fort wayne IN 46818
Manufacturer Contact
zaadvinder sajjan
4545 kroemer road
fort wayne, IN 46818
MDR Report Key24627078
Report Number3004976965-2026-00005
Device Sequence Number8853169
Product Code LXH
UDI-Device Identifier00840096401394
UDI-Public00840096401394
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberT2756
Device Catalogue Number00-1206-090-10
Device Lot Number5737897
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/17/2026
Supplement Date Received by Manufacturer04/15/2026
Initial Report FDA Received Date03/18/2026
Supplement Report FDA Received Date04/16/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/23/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-