ETHICON ENDO-SURGERY, LLC. PWRD 31MM CURVED CIRCULAR, 18CM SHAFT; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number CDH31P
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Failure to Anastomose (1028)
Date of Event 01/01/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Date sent: 3/18/2026.D4: batch # unk.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.D4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.An analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Atempts are being made to obtain the following information.To date, a response has not been received.Should additional information be obtained, a supplemental report will be submitted.What were the indications for surgery? what surgical procedure was performed? did the patient receive any preoperative chemotherapy or radiation? were there any issues experienced with the device in the initial surgical procedure? what healthcare professional fired the device and what is his/her experience with the device? where in the green gap setting scale was the indicator located prior to firing (low-b, middle-b, or high-b)? did the healthcare professional wait 15 seconds after closing the device and then retighten prior to firing? were there any issues with device use/firing? what confirmation was received that the device completed the firing sequence? was the green checkmark visible at the end of the firing? how many counter-clockwise revolutions of the adjusting knob were used to open the device? was there any difficulty removing the device? was a complete transection of the white breakaway washer visually confirmed? were the donuts inspected? if so, please describe.Were there any issues noted with staple formation? if so, please describe the shape and location.Was a leak test performed? if so, what type and what was the result? was the staple line visualized endoscopically during the initial surgical procedure? how many days postoperative did the leak occur? how was the leak identified? what was observed at the site of the leak upon reoperation? how was the leak addressed? what is the current status of the patient? is the surgeon interested in speaking with ethicon medical and engineering staff? this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during a robotic rectum resection procedure, noticed an increase in anastomotic leakage in rectum resections when switching from medtronic circular manual stapler 28 mm to echelon circular powered stapler 31 mm.No further details known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PWRD 31MM CURVED CIRCULAR, 18CM SHAFT
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
*  
3035526892
MDR Report Key24627287
Report Number3005075853-2026-02083
Device Sequence Number8698481
Product Code GDW
UDI-Device Identifier10705036015390
UDI-Public10705036015390
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K163523
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCDH31P
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/19/2026
Initial Report FDA Received Date03/18/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-