| Model Number |
102878 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/19/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A patient implanted with an evoke spinal cord stimulation (scs) system underwent a routine device check prior to a magnetic resonance imagery (mri) scan for an unrelated condition.Disconnected electrodes were identified during an impedance check.A lead revision was subsequently performed to replace the leads and restore mri compatibility.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation of this event is in progress.Once the investigation is complete, a supplemental report will be submitted.Both leads are from the same lot.
|
| |
|