• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL WECK HOL ML 5MM ENDO APPLIER; APPLIER, SURGICAL, CLIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL WECK HOL ML 5MM ENDO APPLIER; APPLIER, SURGICAL, CLIP Back to Search Results
Catalog Number 544965
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/20/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that "a popped rivet was identified during repair at msi." no patient involvement.
 
Additional Manufacturer Narrative
(b)(4).The sample was returned to the manufacturer.The manufacturer reported: "per customer provided information the dhr for the alleged defective medical device was reviewed and found complete without any irregularities.The alleged defective instrument was produced at the tecomet, inc.Kenosha, wl facility as part of a 50 pc.Lot in november of 2024 from lot number: 06e2467427.The investigation determined that the manufacturing processes adhered to approved specifications, utilizing the correct materials and components.A thorough visual and functional inspection was performed.The inspection revealed that the jaws were completely missing and the outer tube was bent.Further evaluation determined that the rivet had dislodged, which likely resulted in the loss of the jaws and deformation of the outer tube, rendering the device nonfunctional.While the exact cause of the missing jaws and the rivet dislodgement could not be conclusively identified, excessive force during use at the end user's facility is suspected.End of life for surgical instruments is normally determined by wear and damage due to the intended use or misuse.Even with proper handling, correct care, and maintenance; reusable instruments should not be expected to last indefinitely and should be replaced at any sign of wear and/or damage.Notably, all 50 instruments from this lot underwent 100% visual inspection and functional testing prior to shipment, in accordance with standard operating procedures for this product line.Due to these findings, no further actions will be taken in response to this complaint, and this record will be deemed closed." teleflex will continue to monitor and trend on complaints of this nature.
 
Event or Problem Description
It was reported that "a popped rivet was identified during repair at msi." no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WECK HOL ML 5MM ENDO APPLIER
Common Device Name
APPLIER, SURGICAL, CLIP
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
3015 carrington mill blvd
morrisville NC 27560
Manufacturer Contact
azhwan barazenda
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key24628314
Report Number3011137372-2026-00063
Device Sequence Number18441917
Product Code GDO
UDI-Device Identifier44026704727481
UDI-Public(01)44026704727481(11)240531(17)001231(10)06E2467427(21)
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number544965
Device Lot Number06E2467427
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2026
Supplement Date Received by Manufacturer03/23/2026
Initial Report FDA Received Date03/18/2026
Supplement Report FDA Received Date03/26/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/31/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
N/A.; N/A.
Patient SexUnknown
-
-